NI 9219 Calibration Procedure
24
ni.com
•
Connect the LO lead of DMM2 as close as possible to the same point that the LO lead of the
NI 9219 connects to the resistor network.
Figure 16.
Full-Bridge Accuracy Verification Connections to the NI 9219 (Configuration 2)
Figure 17.
Full-Bridge Accuracy Verification Connections to the NI 9219 (Configuration 3)
9.
Configure DMM1 for a voltage measurement in the 10 V range.
10. Enable Auto Zero on DMM1.
HI
LO
DMM2
HI
LO
DMM1
V+
V–
Source
10 k
Ω
± 1%
10 k
Ω
± 1%
NI 9219
EX+ (3)
HI (4)
LO (6)
EX– (5)
Ch
x
R
TestPoint
HI
V
LO
DMM2
HI
V
LO
DMM1
V+
V–
Source
10 k
Ω
± 1%
10 k
Ω
± 1%
NI 9219
EX+ (3)
HI (4)
LO (6)
EX– (5)
Ch
x
R
TestPoint