NI 9219 Calibration Procedure
36
ni.com
12. Acquire a 4-wire resistance reading with the NI 9219 using the NI 9219 instance of the DAQmx
Adjust C Series Calibration VI.
13. Reconnect the DMM as shown in Figure 7.
14. Repeat steps 6 through 13 for every value in the array.
15. Compute and save the adjustment to the EEPROM using the DAQmx Close External
Calibration VI. This VI also saves the date, time, and temperature of the adjustment to the
EEPROM.
16. Repeat steps 2 through 15 for each 4-wire resistance mode on the NI 9219.
17. Disconnect the DMM and the resistor from the device.
18. Repeat steps 1 through 17 for each channel on the NI 9219.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust9219Cal
with the following
parameters:
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i
x
*
value
: The DMM reading.
*
x
refers to the channel number
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_Commit