National Instruments Corporation
33
NI 9219 Calibration Procedure
3.
Place the NI 9219 channel in resistance mode using the NI 9219 instance of the DAQmx Setup
C Series Calibration VI.
4.
Input the external temperature using the DAQmx Set Temperature C Series Calibration VI.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup9219Cal
with the following
parameters:
rangeMin
:
0
rangeMax
:
1050, 10500
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i
x
*
measType
:
DAQmx_Val_Resistance
bridgeConfig
:
DAQmx_Val_NoBridge
*
x
refers to the channel number
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCSeriesSetCalTemp
with the
following parameters:
calHandle
:
calHandle
temperature
: The external temperature value in
degrees Celsius.