Remote control
7-9
?TGF
This query command
#?TGF*
sends back a string expressing the tuning frequency,
in exponential notation, for the internal CW signal generator. Unit is fixed MHz.
Example of reply:
TGF = 1.07
which means that the tuning frequency for
CW generator
(when it is not tracking)
is 15 MHz
?TGL
This query command
#?TGL*
sends back a string expressing the output level of the
internal CW signal generator. Unit is fixed dBµV.
Example of reply:
TGL = 90.0
which means that the output level of the CW
generator 90.0 dBµV
?TGS
This query command
#?TGS*
sends back a string expressing the built-in generator
status. The reply is either:
•
TGS=Off
(Built-in generator is disabled)
•
TGS=On
(Built-in generator is active)
?TGT
This query command
#?TGT*
sends back a string expressing the Tracking status
of built-in CW signal generator. The reply is either:
•
TGT=Off
(Built-in generator is no tracking but is tuned on a fixed
frequency, set by STGF)
•
TGT =On
(Built-In generator is tracking the same frequency tuned on the
receiver. Available in Manual Mode only.)
?UHT
This query command
#?UHT*
sends back a string expressing the actual hold time
currently or lastly used
Example of reply:
UHT=1.9ms
?UPP
This query command
#?UPP*
sends back a string expressing the status of input
PIN of the User Port.
Example of reply
: UPP= 1
means that user port pin IN0 is high level
Summary of Contents for PMM ER9000
Page 46: ...3 20 Operating Instructions This page has been left blank intentionally...
Page 50: ...4 4 Applications Fig 4 2 ER9000 with external L2 16B LISN...
Page 52: ...4 6 Applications This page has been left blank intentionally...
Page 66: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 96: ...7 30 Remote control This page has been left blank intentionally...
Page 106: ...Suggerimenti Commenti Note Suggestions Comments Note...