8-6
Click Mode Operating Instructions
8.6 Report
After a Click test the PMM ER9000 reports all the relevant data.
Due to the complexity of the test and to the various conditions that may
happen during the measurement, several reports may be issued.
8.6.1 Fail during
determination of
the click rate N
In case a test failed condition is recorded during the determination of the
click rate N during the first Run, the report will show the situation. In this
case there is no need to go on testing that specific frequency/frequencies.
8.6.2 Report after a
successful test with
less than 5
instantaneous
switching at one
frequency
If the determination of the click rate N shows that at a given frequency there
are less than 5 instantaneous switching operations, there is no need of any
further test at that frequency or frequencies (the investigation frequencies
are 150 kHz for the interval from 150 to 500 kHz, and 500 kHz for the
interval from 500 kHz to 30 MHz).
8.7 Setup
Due to the complexity of the discontinuous disturbance test, this panel is
rather complex, as many parameters shall be defined.
However, the PMM ER9000 has been designed to be extremely user-
friendly and to take care in a fully automatic way of all the necessary steps
of the test.
Due to the fact that the Standard fixes the frequencies at which the
test shall be done, these frequencies have been uniquely assigned by
factory and the User cannot modify them; nevertheless it is possible
to select an arbitrary Frequency Value for "Idle Mode" instead, useful
for debugging purposes.
From the main Click Analysis window, select START and the main Setup
panel will be displayed.
The User can navigate through the different parameters to set the desired
values.
8.7.1 External
attenuator
In case an external attenuator is used to protect the input of the analyzer,
the value of this attenuator can be entered, in order to have the PMM
ER9000 taking into account the value in all calculations of measured levels.
The way to enter the value of the attenuator is shown in the CA0010 User’s
Manual.
8.7.2 Limit
Three most used limits are built-in into the analyzer: CISPR 14, CISPR22
and CISPR11.
Either one can be selected simply by the drop down menu.
The limit will be loaded and all the calculations will be done with reference
to the selected limit:
The Reference Standard for Click tests is always EN 55014-1 (or
equivalent CISPR 14-1), which defines 4 different Limit Curves for Q-Peak
according to Table 1, with following cross-correlation as for names
assigned on ER9000 :
014-1 --> Limit in Column 2
014-2 --> Limit in Column 6
014-3 --> Limit in Column 8
014-4 --> Limit in Column 10
Limit in Column 4 of Tab.1 is not used for Click test, as referred to
Load and Auxiliary terminals.
Summary of Contents for PMM ER9000
Page 46: ...3 20 Operating Instructions This page has been left blank intentionally...
Page 50: ...4 4 Applications Fig 4 2 ER9000 with external L2 16B LISN...
Page 52: ...4 6 Applications This page has been left blank intentionally...
Page 66: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 96: ...7 30 Remote control This page has been left blank intentionally...
Page 106: ...Suggerimenti Commenti Note Suggestions Comments Note...