1-4
General Information
Scan time
A band (9 kHz to 150 kHz)
200 Hz RBW
B band (150 kHz to 30 MHz)
9 kHz RBW
C band (30 MHz to 300 MHz)
120 kHz RBW
D band (300 MHz to 1 GHz)
120 kHz RBW
E band (1 GHz to 3 GHz)
1 MHz RBW
SWEEP MODE
ANALYZER MODE
(CISPR: preselector ON, QP detector)
(preselector OFF, Peak detector, Hold time lowest)
< 2 s (Hold time 1 s)
< 50 ms (Ht 27 ms)
< 3 s (Hold time 2 s)
< 3 s (Hold time 1 s)
< 10 ms (Ht 525 us)
< 5 s (Hold time 2 s)
< 20 s (Hold time 1 s)
< 100 ms (Ht 32 us)
< 40 s (Hold time 2 s)
< 40 s (Hold time 1 s)
< 500 ms (Hold time 32 us)
< 80 s (Hold time 2 s)
< 160 s (Hold time 1 s)
< 400 ms (Hold time 4 us)
< 320 s (Hold time 2 s)
Level measuring time
(Hold time)
CISPR 16-1-1 as default.
2 us to 120 s
Measurement accuracy
S/N > 20 dB
10 Hz to 9 kHz
± 1.0 dB (typ.)
9 kHz to 30 MHz
± 0.8 dB
30 MHz to 1 GHz
± 1.0 dB
1 to 3 GHz
± 1.5 dB
Recommended calibration interval
24 months
Stand-alone main measure functions
Or with free system SW PMM Emission Suite
Manual, spectrum analyser and sweep modes.
Waterfall
Standard and user definable limits
Conversion and correction factors
Control of DDA (Click) analyser, LISNs and other accessories.
Auto diagnosis
Auto calibration
Test reporting
Units
(80 to 200 dB selectable dynamic)
dBm, dBµV, dBµA, dBpW, dBµV/m, dBµA/m, dBpT
Demodulation
AM – FM Internal loudspeaker
RF output
Tracking (manual mode only) & CW generator
Frequency range
Level range
Level accuracy
CISPR pulse generator
Pulse repetition frequency
PRF accuracy
Z
out
50 Ω, BNC female
10 Hz to 30 MHz
60 to 90 dBµV (0.1 dB step)
0.5 dB
1 to 1000 Hz
0.1%
Autocalibration
Internal reference source
Display
Color 6.2” TFT LCD touch panel
I/O Interface
RS-232 DB9, USB 2.0 type B, User port DB15 (Drives PMM LISNs and accessories)
USB type A for memory sticks, Serial Optical interface RP02, Ethernet 10/100 BaseT RJ45
Operating temperature
-5° to 45°C
Power supply
10 - 15 Vdc, 2.5 A with AC universal adapter/charger
Li-ion rechargeable & replaceable battery (BP-02)
Dimensions
235x105x300 mm
Weight
5.0 kg
Summary of Contents for PMM ER9000
Page 46: ...3 20 Operating Instructions This page has been left blank intentionally...
Page 50: ...4 4 Applications Fig 4 2 ER9000 with external L2 16B LISN...
Page 52: ...4 6 Applications This page has been left blank intentionally...
Page 66: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 96: ...7 30 Remote control This page has been left blank intentionally...