Analyzer Mode Operating Instructions
5-5
When clicked, the Custom soft button enables a set of parameters that can
be edited as desired, within the limits of the main specifications.
Fig. 5-3
Custom signals parameters
In the Data Test frame, for the custom disturbance generation, it is possible
to set the duration of up to 4 events, each one with its amplitude, and a
different separation interval between them.
The Repetition box ca be used to set a time interval to program an automatic
N times redo of the events sequence.
Set the duration to 0 ms for each unwanted event.
In the example above, the disturbance generated will be as following
described:
-
the carrier frequency will be 500 kHz (Frequency tab),
-
external correction factor (for cables, attenuators etc) will be zero,
-
wideband CISPR pulses are switched OFF,
- at first the carrier will be ON for 30 ms, with an amplitude of
66.3 dBµV (E1),
-
then will be OFF for 180 ms (Separation),
-
again ON for 30 ms, with an amplitude of 66.3 dBµV (E2),
-
then will be OFF for 100 ms (Separation),
-
again ON for 30 ms, with an amplitude of 46.3 dBµV (E3),
-
OFF for 100 ms (Separation),
-
ON for 30 ms, with an amplitude of 46.3 dBµV (E4),
- The sequence above will be repeated every 3 seconds, for 10 times
(Repetition)
Summary of Contents for PMM CA0010
Page 8: ...VIII Safety considerations This page has been left blank intentionally...
Page 18: ...1 10 General Information This page has been left blank intentionally...
Page 24: ...2 6 Installation This page has been left blank intentionally...
Page 49: ...Sweep Mode Operating Instructions 4 3 Fig 4 4 Installation folder Fig 4 5 Ready to install...
Page 72: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 74: ...Suggerimenti Commenti Note Suggestions Comments Note...