3-12
Setup and Panel Instructions
3.8.2 Limit
Fig. 3-5
Limit
Three most used limits are built-in into the analyzer: CISPR 14, CISPR22
and CISPR11.
Either one can be selected simply by pressing the relevant key.
The limit will be loaded and all the calculations will be done with reference
to the selected limit:
The Reference Standard for Click tests is always EN 55014-1 (or
equivalent CISPR 14-1), which defines 4 different Limit Curves for Q-Peak
according to Table 1, with following cross-correlation as for names
assigned on 9010 :
014-1 --> Limit in Column 2
014-2 --> Limit in Column 6
014-3 --> Limit in Column 8
014-4 --> Limit in Column 10
Limit in Column 4 of Tab.1 is not used for Click test, as referred to
Load and Auxiliary terminals.