4-8
Sweep Mode Operating Instructions
4.4.1 Setup
When the software button is clicked, the Click Setup window pops up.
Here it is possible to:
•
set the value in dB of a possible external attenuator,
•
choose the standard limit,
•
set if rate calculation will be based on RF signal measurement
(Clicks) or on switching operations (current consumption threshold),
•
set the f factor, when needed, as given in specific CISPR 14 Annex,
•
choose if terminate current step or go on anyway, when the test has
failed,
•
choose if terminate current run manually during acquisition, or
alternatively when 40 clicks are reached. In any case the step
terminates when the expected observation time is elapsed,
•
in the Line frame it is possible to select the desired line of a PMM
LISN, if used in the test setup. PMM LISNs can be automatically
controlled by the 9010F receiver via the User Port connection,
•
set the maximum observation time,
•
set Idle threshold level in dBµV. This is the threshold level, for clicks
detection, when a standard limit is not selected,
•
choose if Single Run smart function will be adopted or not,
•
choose if use Exception 4, as provided by the standard,
•
select if rate calculation is based on the current (when this manual
was edited for the first time) standard edition (2 channels, 2
quartiles), or taking into account 2 channels, and then 4 quartiles, or
even considering a possible future regulation with 4 channels and 4
quartiles,
•
Set Click Setup soft button sends the settings to the analyzer when
clicked,
•
Default button retrieves a base standard set of parameters.
•
digit the Name of the measurement to be executed and,
consequently, archived,
•
Start soft button runs the acquisition after being clicked,
•
Idle soft button is used to enter the Idle mode, for EUT initial
evaluation, debug and testing purposes,
•
Exit soft button to close the Start window without running a new
measurement.
Summary of Contents for PMM CA0010
Page 8: ...VIII Safety considerations This page has been left blank intentionally...
Page 18: ...1 10 General Information This page has been left blank intentionally...
Page 24: ...2 6 Installation This page has been left blank intentionally...
Page 49: ...Sweep Mode Operating Instructions 4 3 Fig 4 4 Installation folder Fig 4 5 Ready to install...
Page 72: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 74: ...Suggerimenti Commenti Note Suggestions Comments Note...