When using a PMM LISN, it is possible to control the lines from the receiver,
and automatically switch between them by using a special cable between
the User Port of the receiver and the LISN remote control input.
Shift the back power switch to the OFF position to switch the unit OFF.
Fig. 2-1
PMM CA0010 click analysis setup.
2.11 Using an Artificial
Mains Network
(AMN or LISN)
When the PMM CA0010 analyzer is connected to an external LISN to
perform conducted interference measurements of the EUT, the RF output of
the LISN shall be connected to the RF input of the analyzer. In order to
switch automatically between the lines of the LISN, it shall be connected to
the User Port of the receiver thanks to the dedicated cable supplied with the
LISN.
Using the provided PMM Click Analysis software, it is possible to carry out
an automatic measurement on all mains lines.
In order to avoid the unwanted tripping of the protection devices, an
insulation transformer shall always be used between the mains supply
and a LISN.
When using an external Artificial Mains Network, to reduce the
probability of damages caused by transient voltage pulses, always
disconnect the PMM CA0010 RF input before switching Equipment
Under Test supply ON or OFF.
Click Meter Reports can be remotely downloaded to the PC both using
the PMM Click Analysis and PMM Emission Suite Software.
Summary of Contents for PMM CA0010
Page 8: ...VIII Safety considerations This page has been left blank intentionally...
Page 18: ...1 10 General Information This page has been left blank intentionally...
Page 24: ...2 6 Installation This page has been left blank intentionally...
Page 49: ...Sweep Mode Operating Instructions 4 3 Fig 4 4 Installation folder Fig 4 5 Ready to install...
Page 72: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 74: ...Suggerimenti Commenti Note Suggestions Comments Note...