![multicomp pro MP720021 Manual Download Page 96](http://html1.mh-extra.com/html/multicomp-pro/mp720021/mp720021_manual_1854297096.webp)
User Manual
91
Source
CH1 to CH4
Display the channel to which the cursor
measurement will be applied.
Perform the following operation steps for the time and voltage cursor measurement
of the channel CH1:
1.
Push
Cursor
to display the cursor menu.
2.
Select
Source
in the bottom menu, select
CH1
in the right menu.
3.
Select the first menu item in the bottom menu, the
Type
menu will display at
the right of the screen. In the right menu, select
Time&Voltage
for Type,
two blue dotted lines displayed along the horizontal direction of the screen,
two blue dotted lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen shows the cursor
readout.
4.
In the bottom menu, select
Line Type
as
Time
to make the vertical cursors
active. If the
Line
in the bottom menu is select as
a
, turn the
M
rotary control
to move line a to the right or left. If
b
is selected, turn the
M
rotary control to
move line b.
5.
In the bottom menu, select
Line Type
as
Voltage
to make the horizontal
cursors active. Select
Line
in the bottom menu as
a
or
b
, turn the
M
rotary
control to move it.
6.
Push the
Horizontal
HOR
button to enter wave zoom mode. In the bottom
cursor menu, select
Window
as
Main
or
Extension
to make the cursors
shown in the main window or zoom window.
Figure 4-27 Time&Voltage Cursor Measurement
Summary of Contents for MP720021
Page 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Page 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Page 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Page 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Page 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Page 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...