User Manual
83
Rule
Source
Horizontal
Vertical
Create
Select the source as CH1, CH2, CH3 or CH4
Change the Horizontal tolerance value by turning the
M
rotary control
Change the Vertical tolerance value by turning the
M
rotary control
Use the rule set as testing rule
SaveRule
Number
Save
Load
Select any one from Rule1 - Rule8 as your rule name
Select
Save
to save the rule
Load some rule as the testing rule
The description of
Pass/fail
Menu
is shown as the follows:
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits of
the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass signal by
built-in and configurable output port. To run the test, read the following steps:
1.
Push the
Utility
button, select
Function
in the bottom menu, select
Pass/fail
in the
left menu.
2.
Enable switch on
: Select
Operate
in the bottom menu, select
Enable
in the right
menu as
ON
.
3.
Create rule
: Select
Rule
in the bottom menu. Select
Source
in the right menu,
select the source in the left menu. Set
Horizontal
tolerance and
Vertical
tolerance
in the right menu. Select
Create
in the right menu to create the rule.
4.
Set output type
: Select
Output
in the bottom menu to enter output option setting.
Choose any one or two of the options "
Pass
", "
Fail
" or "
Beep
". "
Pass
" and "
Fail
"
are mutually exclusive options, which could not be chosen simultaneously. "
Stop
"
means stop once the condition satisfies your setting.
5.
Begin to test:
Select
Operate
in the bottom menu, select
Operate
in the right
menu as
Start
, the test will begin.
6.
Save rule:
Select
SaveRule
in the bottom menu. Select the save location in the
left menu, and then select
Save
in the right menu to save the rules, which could be
called up at once when need. Select
Load
to call up the rule saved.
Note:
1.
When Pass/Fail is ON, if XY or FFT is ready to run, then Pass/Fail will be closed;
under the mode of XY or FFT, Pass/Fail is unable.
2.
Under the mode of Factory, Auto Scale and Auto Set, Pass/Fail will be closed.
3.
When no save setting left in the rule save, tip will be given to show "NO RULE
SAVED".
4.
Under the status of stop, data comparing will stop, and when it goes on running,
the number of Pass/Fail will increase from the former number, not from zero.
5.
When the waveform playback mode is on, Pass/Fail is used to test the
played-back waveform specially.
●
Output
Push the
Utility
button, select
Function
in the bottom menu, select
Output
in the left
menu.
Summary of Contents for MP720021
Page 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Page 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Page 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Page 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Page 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Page 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...