The automatic measurement can not be performed in the following situation: 1) On
the saved waveform. 2) On Waveform Math waveform. 3) On the Video trigger
mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1.
Push the
Measure
front panel button to show the Measure menu.
2.
Select
Add
in the bottom menu.
3.
In the left Type menu, turn the
M
rotary control to select
Period
.
4.
In the right menu, select
CH1
in the
Source
menu item.
5.
In the right menu, select
Add
. The period type is added.
6.
In the left Type menu, turn the
M
rotary control to select
Frequency
.
7.
In the right menu, select
CH1
in the
Source
menu item.
8.
In the right menu, select
Add
. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically
(see
Figure 4-24 Automatic measurement
The automatic measurement of voltage parameters
The oscilloscopes provide automatic voltage measurements including Mean, PK-PK,
RMS, Max, Min, Vtop, Vbase, Vamp, OverShoot, PreShoot, Cycle RMS, and
Cursor RMS.
below shows a pulse with some of the voltage
measurement points.
Summary of Contents for MP720021
Page 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Page 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Page 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Page 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Page 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Page 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...