70
Maintenance and Service
ALTAIR io360
US
(10) Install screws with a torque of 3 in-lb +/- 0.5 in-lb (0.3Nm +/- 0.1Nm).
(11) Install new batteries into the cavity oriented such that the battery
matches the orientation marked in the cavity.
See reference and warnings in section 8.8 "Batteries".
(12) Assembly front and back housing. Install screws until resistance
starts.
(13) Finish tightening screws in the pattern as shown.
Tighten screws to a torque of 10 in-lb +/- 1 in-lb (1.1 Nm +/- 0.1 Nm)
Reminder: The detector must pass calibration before continuing use.