23
Hardware Setup
ALTAIR io360
US
To ensure the ingress protection rating IP67:
• all seals must be undamaged and correctly fitted
• all screws of the housing / housing cover must be tightened with a torque of 13 in-lb +/- 1 in-lb (1.5 Nm
+/- 0.1 Nm)
• only use MSA supplied cables
• all connectors must be hand tightened
• If the ethernet connection is not is use, it must be sealed with MSA supplied blind plug.
IP ratings do not imply that the equipment will detect gas during and after exposure to these conditions.
3.5
HUB Assembly
Only perform assembly with the HUB powered off.
(1) Hand-screw antenna.
(2) Open housing by opening the joints and removing the two lid screws
as shown.
Optional: HUB cellular communication
Supported with most major cellular carriers. Further verification may be
required in your area.
Acceptable SIM Cards Types:
• Micro SIM
• Nano SIM with Nano SIM to Micro SIM adapter
(3) Insert SIM card.
NOTE:
SIM card is not provided.