49
Calibration and Bump
ALTAIR io360
US
(4) Tap the Bump RFID tag in Calibration Cap to the button.
Once prompted, a detector will display 'Install Cap', at which time
Install the calibration cap.
(5) Snap the Calibration Cap onto the detector.
(6) Insert tab on Calibration Cap into slots on detector (3 places).
(7) Press Calibration Cap until it seats onto detector.
a) If needed, connect one end of the tubing to the Calibration Cap.
b) If needed, connect other end of tubing to the cylinder regulator.
(8) When the detector shows blue LEDs and prompts 'Apply Gas', apply
gas.
(9) Open valve on regulator to allow gas to flow to detector.