Software Configuration Setup
ALTAIR io360
US
4.6.3 Internet Connectivity
The HUB can be connected to the internet for Grid connectivity with Ethernet, Wifi, and/or cellular.
The HUB will, in that order, preferrably use Ethernet, Wifi, or cellular for internet connectivity.
4.7
Modifying a Network
Connect the app to the HUB to:
• Add detector(s) to network: max. 32 for 915 Mhz and 16 for 868 Mhz
• Remove detector(s) from network
• Edit network name
• Enable/disable shared alarms
4.7.1 Shared Alarms
A shared alarm is an alarm that, from the alarm source through the network, reaches every device on the
same network.
Shared alarms latch and non-latch based on your alarm settings.
There are two types of shared alarms:
• Network triggered shared alarm
• Manually initiated shared alarm
4.7.2 Evacuation
An evacuation alarm can be initiated manually and communicated through the network at the discretion of
an authorized user.
Alarms are configured as shared alarms by default.
The user can define evacuation alarms to not be shared through the network.
4.7.3 Alarm Settings
Factory default alarm set points are stated in the table below and can be changed within the stated limits.
Use the app to change the alarm settings.
Latching Alarm Configurations
A latching alarm persists until acknowledged by the user, even if the gas reading returns to non-alarm levels.
The factory default settings for alarms are set to non-latching.
Alarm latching behavior can be configured for each alarm type in the app.
Sensor
Unit of
Measure
Low Alarm
Default
High Alarm
Default
Min Low
Alarm
Max High
Alarm
Combustible
% LEL
10
20
5
60
Combustible
% Vol
0.50
1.00
0.25
3.00
CO
PPM
25
100
7
1700
H
2
S
PPM
10
15
5
175
O
2
% Vol
19.5
23.0
5.0
24.0