45
Operation
ALTAIR io360
US
6.7.1 Factory Reset
If the device access password has been lost and no further access is possible, a factory reset can be
performed to reset all values to their defaults.
(1) Power off the HUB.
(2) Unmount the HUB.
a) If necessary, move to a non hazardous location.
(3) Open housing by opening the right joints and unscrew the lid screws
using a Torx T20.
(4) Power the device.
(5) Use a thin object non-conductive (e.g. a pencil) to push the reset
button for 10 seconds. All LEDs switch on and off.