76
Technical Data
ALTAIR io360
US
11
Technical Data
11.1 Detector
Weight
27.2 oz / 771 g (detector with battery and clip)
Dimensions (L x W x H)
5.2 x 5.6 x 2.7 inches/132 x 142 x 69 mm
Alarms
Visible LEDs and 90 dB at 30 cm horn
Display
LCD display
Battery type
Primary Non-rechargeable Lithium battery. Only replace with Tadiran model
number TL-5920 and Tadiran model SL-2770
Detector run time
50-60 days at 25 °C (77 °F) combustible sensor installed 80-85 days at
25 °C (77 °F) toxic/O
2
only.
These conditions can significantly reduce battery run time:
• Operation at the low and/or high temperatures
• Continuously alarmed state with horn sounding and LEDs flashing
• Continuous background level of gas present
Power up time
2 min during start-up procedure
Temperature Range
Normal operating range:
14 to 104 °F (-10 to 40 °C)
Extended operating range:
-4 to 140 °F (-20 to 60 °C)
Intrinsic safety ambient temperature
range (NA, ATEX, IEC):
-40 to +140 °F (-40 to 60 °C)
Storage:
32 to +95 °F (0 to 35 °C)
Humidity Range
15 % - 90 % relative humidity, non-condensing,
5 % - 95 % RH intermittent
Atmospheric pressure range 800 to 1200 mbar
Ingress protection
IP 68 (2 meters for 60 minutes)
IP ratings do not imply that the equipment will detect gas during and after
exposure to these conditions
Measuring methods
Combustible gases: Catalytic sensors
Oxygen: Electrochemical sensor
Toxic gases: Electrochemical sensor
Measuring Range
Combustible:
0-100% LEL
0-5.00% Vol. CH
4
O
2
0-30% Vol.
CO
0-1999 ppm
H
2
S
0-200 ppm