background image

5

must be in their original packaging prior to use in a contaminated environment. 

•   DO NOT replace canister/cartridge(s) in a contaminated area. Be sure to follow applicable 

decontamination procedures. Failure to follow this warning can cause inhalation of 
contaminated air, resulting in serious respiratory injury or death.

•   DO NOT remove respirator until respirator and protective clothing are decontaminated; otherwise, 

exposure to contaminants may result. Follow decontamination and disposal procedures 
established by appropriate authorities.

Failure to follow the above warnings can result in serious personal injury or death.

RESPIRATOR USE LIMITATIONS

The wearer must comply with the following MSA respirator use limitations:
1.   MAXIMUM USE CONCENTRATION — Do not exceed any of the following:
 

a.  10 times the exposure limit for the contaminants present.

 

b.  Immediately dangerous to life or health (IDLH) concentration for any contaminant present.

2.   The limitations outlined in the applicable NIOSH approval.
3.   Replace particulate filter cartridges whenever they are damaged, soiled or causing noticeably 

increased breathing resistance.

4.   Applicable respirator use requirements as specified in the OSHA Respiratory Protection Regulation 

29 CFR Part 1910.134 (or other requirements established by the Regulatory Agency with jurisdiction 
over the wearer). Additional OSHA Regulations may also apply for certain contaminants (See MSA’s 
Response Respirator Selector).

EXPOSURE LIMITS

A listing of acceptable exposure limits from the following sources is provided in MSA’s Response

®

 

Respirator Selector:
 

—   American Conference of Governmental Industrial Hygienists (ACGIH)

 

—   Occupational Safety and Health Administration (OSHA)

 

—   National Institute for Occupational Safety and Health (NIOSH)

 

—   American Industrial Hygiene Association (AIHA)

Contact MSA at 1-800-MSA-HOW2 (1-888- 672-4692) for information.

RESPIRATOR FIT TEST

A qualitative or quantitative respirator fit test must be carried out for each wearer of this respirator to 
determine the amount of protection it will provide. Respirator fit tests are explained fully in ASTM F3387 
- Latest edition, Standard Practices for Respiratory Protection, ASTM International https://www.astm.org/
Standards/F3387.htm? (ASTM F3387 formerly known as ANSI Z88.2)

QUANTITATIVE TEST 

– If a Quantitative Fit Test is used, a fit factor that is at least 100 shall be obtained 

before that respirator is assigned to an individual.

QUALITATIVE TEST 

- If a Qualitative Fit Test is used, only validated protocols are acceptable. The 

individual must pass a test designed to assess a fit factor of at least 100.

TAL 122 (L) Rev. 0 - 10218530

Summary of Contents for Advantage 900

Page 1: ...respirator will perform as designed only if used and maintained according to the instructions Otherwise it could fail to perform as designed and persons who rely on this product could sustain serious...

Page 2: ...tretien de ces appareils de protection respiratoire P NIOSH n value pas l usage des appareils de protection respiratoire comme masques chirurgicaux S Des instructions sp ciales ou critiques et ou des...

Page 3: ...otection against specific contaminants 3 This respirator is not approved for use with chemical or combination cartridges Use only with the particulate cartidges listed in the approvals matrix 4 Do not...

Page 4: ...gh or sneeze and spread virus particles through the air to exposed surfaces which are touched Tight fitting safety goggles or a full facepiece respirator may further help prevent transmission of viral...

Page 5: ...ction Regulation 29 CFR Part 1910 134 or other requirements established by the Regulatory Agency with jurisdiction over the wearer Additional OSHA Regulations may also apply for certain contaminants S...

Page 6: ...ilter by washing it or using compressed air Inspect cartridges for scratches cracks or other damage particularly the sealing bead around the bottom CARTRIDGE ATTACHMENT Carefully attach cartridges to...

Page 7: ...sition the facepiece ensure the push button is completely depressed and repeat test until no leakage is detected If any leakage is detected around the facial seal readjust head harness straps and repe...

Page 8: ...ighten each cartridge by gripping as much of the circumference of the cartridge as possible and then slowly turning the cartridge until tight CLEANING AND DISINFECTING DO NOT use alcohol as a germicid...

Page 9: ...teps below to reassemble 1 Stretch the gasket around the speaking diaphragm The gasket should be installed so that the thin ner rim of the gasket is on the side of the diaphragm with 4 small bumps 2 I...

Page 10: ...distortion of parts which would require replacement MAINTENANCE This respirator must be kept in good condition to function properly When any respirator shows evidence of excessive wear or damage it mu...

Page 11: ...correctamente el uso y el mantenimiento deben realizarse conforme a las instrucciones proporcionadas De lo contrario podr a no ofrecer el rendimiento para el cual est dise ado y ocasionar lesiones gr...

Page 12: ...es appareils de protection respiratoire P NIOSH n value pas l usage des appareils de protection respiratoire comme masques chirurgicaux S Des instructions sp ciales ou critiques et ou des limites d em...

Page 13: ...espec ficos 3 Este respirador no est aprobado para el uso con cartuchos qu micos o combinados selo nicamente con los cartuchos para part culas que figuran en la tabla de aprobaciones 4 No use el prod...

Page 14: ...en todo momento como si estuviera contaminada Un respirador NO previene la exposici n a la gripe o a otros agentes biol gicos suspendidos en el aire de otras maneras como toc ndose al boca la nariz o...

Page 15: ...so omiso de estas advertencias e instrucciones puede comportar lesiones graves o incluso la muerte RESTRICCIONES DE USO DEL RESPIRADOR El usuario debe respetar las siguientes restricciones de uso del...

Page 16: ...o El usuario debe realizar una prueba de ajuste del respirador y respetar todas las advertencias y restricciones dadas Hacer caso omiso de esta advertencia puede comportar lesiones graves o incluso la...

Page 17: ...ebe quedar c moda alrededor de la nariz de la boca y del ment n 4 Tire de la banda de soporte de la cabeza por encima de la cabeza de manera que cubra la coronilla La banda de soporte de la cabeza deb...

Page 18: ...r una de estas pruebas de estanqueidad antes de que pueda usarse No entre en ninguna atm sfera con este respirador a menos de que est seguro de que se cumplen las siguientes condiciones 1 Ha le do com...

Page 19: ...mente se debe desinfectar para eliminar los microorganismos pat genos despu s de cada uso con el limpiador Confidence Plus 2 de MSA EE UU P N 10009971 Canad P N 10032737 Siga las instrucciones del lim...

Page 20: ...llado debe presionarse en la m scara completamente antes de presionar la segunda superficie de sellado Aseg rese de que el empaque quede plano sin enrollarse 3 Conecte el ensamble de sujeci n en la m...

Page 21: ...e las piezas que de no ser as tendr an que reemplazarse MANTENIMIENTO Este respirador debe mantenerse en perfectas condiciones para que pueda funcionar correctamente Cuando un respirador presenta se a...

Page 22: ...22...

Page 23: ...il se doit uniquement s il est utilis et entretenu conform ment aux instructions Autrement il pourrait ne pas donner le rendement pr vu et les personnes qui d pendent de ce produit risqueraient de sub...

Page 24: ...tretien de ces appareils de protection respiratoire P NIOSH n value pas l usage des appareils de protection respiratoire comme masques chirurgicaux S Des instructions sp ciales ou critiques et ou des...

Page 25: ...n est pas homologu pour l utilisation avec des cartouches chimiques ou combin es L utiliser seulement avec les cartouches pour particules qui figurent dans le tableau d homologation 4 Ne pas utiliser...

Page 26: ...grammes de protection respiratoire contacter l OSHA ou visiter le site Web www osha gov Ne PAS retirer le respirateur dans des zones contamin es La surface externe du respirateur DOIT en tout temps tr...

Page 27: ...ments de protection n ont pas t d contamin s autrement une exposition des contaminants pourrait survenir Respecter les proc dures de d contamination et d limination tablies par les autorit s comp tent...

Page 28: ...se l utilisateur des risques de blessures graves ou mortelles MESURES PR LIMINAIRES L UTILISATION Les points d inspection suivants doivent tre v rifi s avant d enfiler le respirateur Tout respirateur...

Page 29: ...rine La bande arqu e du serre t te doit reposer plat sur le cr ne sans tre tordue 3 Tirer sur les sangles de mani re relever le masque jusqu ce qu il couvre le nez et la bouche 4 Raccorder les boucles...

Page 30: ...s du joint facial on doit proc der une inspection plus approfondie et corriger le probl me avant d effectuer un nouvel essai Le respirateur doit r ussir l un des essais d tanch it ci dessus avant d tr...

Page 31: ...ciale doit tre d contamin e enlever d abord les cartouches La pi ce faciale doit d abord tre nettoy e pour liminer la salet et les huiles Puis apr s chaque utilisation on doit la d sinfecter pour limi...

Page 32: ...La premi re surface d tanch it doit d abord tre com pl tement enfonc e dans le masque ensuite on peut y enfoncer la deuxi me surface d tanch it S assurer que le joint est pos plat sans tre tordu 3 Fi...

Page 33: ...emplac es ENTRETIEN Ce respirateur doit rester en bon tat pour pouvoir fonctionner correctement Lorsqu un respirateur montre des signes d usure ou de dommage excessifs il doit tre remplac imm diatemen...

Page 34: ...34...

Page 35: ...35...

Page 36: ...36...

Reviews: