Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 40
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Option Printer Command
Code
Baudrate
0
300
1
600
2
1200
3
2400
4
4800
5
9600
6
19.2K
7
Special
Option Tare Retain Command (OT)
The Retain Tare (OT) command enables and disables the retention of the current tare value in memory for use
after the unit has been powered down and then powered back up. The format for the OT command is: OT(0 or
1), where 0 is disable and 1 is enable. Once enabled, the tare values are retained until this option is disabled.
Disabling the tare retention option resets the tare values for both channels to 0.
Example:
@123OT1
Acknowledge:
@123 Retain tare is on
Display Options Command (D)
The Display options command has 5 subcommands used to set options that effect the front panel display of
the Model 4215-HS.
Display View Command (DV)
The Display View (DV) command is used to return the current values of all the display options. The format is
DV.
Example:
@123DV
Acknowledge:
@123 Filter is Type II Level 2
Channel A shows 4 decimal digits
Channel B shows 4 decimal digits
Channel A counts by 1
Channel B counts by 1
Second line shows limit status