Calibrate Counts-Per-Inch Command (CI)
This command has two subcommands, a view command (CIV) and a set command (CIS).
The Calibration counts per Inch View (CIV) command displays the current setting of the encoder counts per
inch constant.
Example:
@123CIV
Acknowledge:
@123 Encoder is 32000 counts per inch
The Calibration counts per Inch Set (CIS) command allows you to change the current setting of the encoder
counts per inch constant.
Example:
@123CIS64000#
Acknowledge:
@123 Encoder is 64000 counts per inch
Limits Command (L)
This command allows you to view or set the 4 contact closure limit switches on the Model 4215. There are
three subcommands Limit View (LiV), Limit Set (LiS), and Limit Reset (LiR).
Limit View Command (LiV)
The Limit View (LiV) command is used to view the current settings of one of the limits. The format is L(limit
number)V.
Example:
@123L1V
Acknowledge:
@123 Lim 1 NO Enabled Load A Lb Set 20.00 Trip>Set Latch Off Reset 1.000
This line indicates that Limit 1 is a normally open switch (NO) rather than a normally closed switch (NC). The
limit is Enabled rather than Disabled. The signal driving this limit is Load in Lb. The limit will be turned on when
the load exceeds 20.00 Lb. It is not a Latching limit (latching requires a manual reset). The limit will automati
-
cally reset when the load is less than 1.00 Lb.