Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Setup Mode
Setup mode is used to change the setup of the Model 4215-HS. To enter setup mode press the two end but
-
tons at the same time ( “<“ and “>“ ). The row of four buttons will now function as indicated by the lower set of
labels (Plus, Minus, Enter, and Escape).
If the optional keylock is in the locked position you will see “>Setup is Locked” when you try to enter setup
mode. This prevents accidental or unauthorized changes to the setup of the unit.
To exit from setup mode and return to run mode press the escape button (Esc). When you are entering numer
-
ical data or selecting options from a sub-menu the escape button will back you out without making any chang
-
es.
The main menu in setup mode contains the following 6 entries which will be described in detail in the follow
-
ing sections. Note that all main menu items begin with the “>“ symbol and that all submenu items begin with
the “>>“ symbol.
• > User Data Entry -- allows you to set any common math options such as the base area used to display load
in PSI or the base length used to display position in percent.
• > Analog Output -- is used to select the analog channel, source, and scaling for the analog output channels.
• > Sensor Select -- lets you manually set which load cell calibration data is used. If you use auto-id load cells
you will never need this menu.
• > Calibration -- provides access to calibration for load cells, extension sensor, or optical encoders.
• > Limits -- allows you to set the source and options for the 4 contact closure switches.
• > System Options -- contains entries to set the printer baud rate, enable or disable automatic sensor identifi
-
cation, enable or disable automatic load zeroing, and to set the serial communications address, baud rate, and
line feed options.
• > Display Options -- lets you choose the filtering that is applied to the load measurements, the number of
decimal digits that are shown on the display and on the print, the count by values for channels A and B (1, 2,
5, 10, or 20), and whether the second line of the display shows the limit switch status, a blank line, another
display, or text.
The Plus or Right button ( “+” or “>“ ) will scroll forward through this menu. The Minus or Left button ( “-” or
“<“ ) will scroll backward through this menu. The Enter button selects the currently displayed menu item. The
Esc button will return you to Run Mode.
Entering Numerical Data
At times you will need to enter numerical data into the Model 4215-HS. The method is consistent in all cases. A
number is presented on the left side of the display with an underline cursor.