Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 26
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Rs232 Setup Mode Commands
The commands in this section allow you to change the setup of the Model 4215-HS. You will notice that these
commands form a menu which exactly duplicates the Setup Mode menu used during setup of the Model 4215-
HS from the front panel.
If the optional Keylock is installed and locked you will not be to alter the setup until the unit has been un
-
locked.
The acknowledge will indicate this fact:
@123 Setup is Locked - use Keylock to unlock
User Data Entry Command (U)
This command is used to view or set the options in the user data entry menu, namely the base area (used to
calculate load in PSI, or MPa) and base length (used to calculate position in %). This command has three sub
-
commands: User data View (UV), User data Area (UA), and User data Length (UL).
User Data View Command (UV)
The User data View (UV) command has the format UV.
Example:
@123UV
Acknowledge:
@123 Base Area Ch A is 1.00000 sq-in
Base Area Ch B is 0.50000 sq-in
Base Length is 3.0000 in
User Data Area Command (UA)
The User data Area (UA) command is used to set the base area. The format is UA(channel)(number)#.
Example:
@123UAA1.0025#
Acknowledge:
@123 Base Area Ch A is 1.0025 sq-in
User Data Length Command (UL)
The User data Length (UL) command is used to set the base length. The format is UL(number)#
Example:
@123UL2.5#
Acknowledge:
@123 Base Length is 2.5000 in