Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 38
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Option View Command (OV)
The Option View (OV) command returns a list of the current settings of all the system options. The format is
OV.
Example:
@123OV
Acknowledge:
@123 Printer Baud Rate is 9600
Auto Identify is off
Auto Zero Channel A is off
Auto Zero Channel B is off
Com Address is 123
Com Baud Rate is 4800
Com Line Feed is off
Retain Tare is off
Option Printer Baudrate Command (OP)
The Option Printer baudrate (OP) command is used to select the printer baudrate. The format is OP(code)
where the code is from the table shown at below:
Example:
@123OP9
Acknowledge:
@123 Printer Baud Rate is 230.4K
Option Auto-Identify Command (OI)
The Option auto-Identify (OI) command is used to turn auto-identification of load cells on or off. The format is
OI(0 or 1), where 0 turns auto-id off, and 1 turns auto-id on.
Example:
@123OI0
Acknowledge:
@123 Auto Identify is off
Option Printer Command
Code
Baudrate
4
4800
5
9600
6
19.2K
8
57.6K
9
230.4K