www.miinet.com
Moore Industries-International, Inc.
- 7 -
User’s Manual
224-790-00K
May 2018
Programmable Limit Alarm Trips
SPA
2
SECTION 1
SPA
2
TPRG Quickstart Guide
Use the front pushbuttons to quickly and easily
set-up the SPA
2
for your application. After
programming your alarm using the diagram
below, install the unit into your application
using the connection diagrams and terminal
designation table located in this manual.
Front Panel
Pushbuttons
Use the UP and DOWN pushbuttons to scroll through menus and
sub-menus. Use the SELECT pushbutton to access menus and/or
make a choice.
Trim Analog Output:
Connect the SPA
2
into a calibration
set-up as shown in the manual and
trim the analog output.
(Requires -AO Option)
Password:
View or change password.
Configure Options:
Display mode, decimal places,
50/60Hz filter, start delay, display test
and factory reset.
CONFG
OPTNS
TRIM
AOUT
Apply Input:
Using calibration equipment,capture
the zero and full scalev alues.
APPLY
INPUT
CONFG
PASS
Exit Configuration Menu:
Return to process display.
CONFG
EXIT
D
D
U
CONFG
ALARM
D
SCALE
AOUT
Scale Analog Output:
Set the values for output at
zero and full input
(Requires -AO Option)
D
U
D
CONFG
AOUT
Configure Analog Output:
Choose damping value
and fail mode.
(Requires -AO Option)
TRIM
INPUT
Trim Input:
Match the actual reading of your
SPA
2
to a calibrated or known
value.
D
U
U
D
U
U
U
D
U
ENTER
PASS
Enter the password
code to enable
settings changes.
(Default = 55)
(
If the incorrect password is entered,
access to the configuration menus is denied.)
Configure Input:
Select Sensor Type
Select Input Zero/Full
Security Jumper
installed in Full
Acess position.
Security Jumper
installed in Password
Required position.
CONFG
SENSR
SELECT
DOWN
UP
SCALE
INPUT
D
U
D
U
D
Scale Input:
Set the values to be displayed
at zero and full scale without
calibration equipment.
System Information:
View information such as
serial number, Firmware Version,
Hardware Version, Input Model,
and Output Model
SYS
INFO
D
Configure Alarm:
Configure alarm types & settings.
Default/Factory Configuration
The following are the default factory settings for your
unit.
Input 4W RTD, 0-100°C
Display Normal Mode, PV, PT 3850
60Hz Filter
Broken wire enabled
Running Average Filter set to 4
All alarms set to Trip High at 50°C with Deadband set to 0
All alarms range set to 0-100
All alarms have OOR & Sensor Failure disabled
All alarms are configured as Fail Safe, latching disabled, 0 delay
AO (if fitted) : Current 4-20mA, fail high, hold duration of 1 second