APPENDIX E – ALARMS
77
which is automatically transformed to the message:
( A l a r m M e s s a g e )
R E S E T S Y S T E M
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Perform a RESET (described below) and then turn-off the device and check the alarm.
NOTE
A
LARMS THAT INTERVENE OUTSIDE OF A PROGRAM DO NOT PRODUCE A PRINTED
REPORT
.
RESETTING THE
SYSTEM
The system is RESET in two alternative ways, depending on the alarm that occurred (see the
Alarm Code List further below in this appendix):
1. Press
the PROGRAM SELECTION key for about 3 seconds.
A beep confirms the RESET;
WARNING
NEVER TURN THE DEVICE OFF BEFORE EXECUTING A RESET.
2. Turn-off
the
device and then power-on using the main switch.
Upon power-up, the sterilizer will perform its normal initial test.
After RESET, and any technical intervention necessary to eliminate the fault, the device will go
to STAND-BY mode, ready to execute a new program.
>3s