9. RUNNING THE PROGRAM
44
Drying
Ventilation
Leveling to the
atmospheric pressure
Completion
of the cycle
After the steam under pressure is released, its forced removal begins with the vacuum pump
(DRYING): for this purpose, low pressure is created in the sterilization chamber to facilitate the
evaporation of the steam and its consequent elimination. As a function of the type of drying
set, one of the following screens will appear:
1
1 3 4 ° C P O R O U S
D R Y I N G ( N O R )
1 0 1 . 1 ° C
0 . 0 0 b a r 1 8 : 5 1
1
1 3 4 ° C P O R O U S
D R Y I N G ( I N T )
1 0 1 . 1 ° C
0 . 0 0 b a r 1 8 : 5 1
1
1 3 4 ° C P O R O U S
D R Y I N G ( + X X )
1 0 1 . 1 ° C
0 . 0 0 b a r 1 8 : 5 1
When the drying phase is finished, it is followed by a VENTILATION phase in which fresh
sterile air is injected, while maintaining a vacuum in the chamber, to eliminate condensate and
cool the load.
1
1 3 4 ° C P O R O U S
V E N T I L A T I O N
8 4 . 4 ° C
- 0 . 7 7 b a r 2 6 : 5 1
At the end of the ventilation phase, the chamber is brought back to atmospheric pressure
(LEVELLING) by injecting sterile outside air to allow the opening of the door and the retrieval
of the load.
1
1 3 4 ° C P O R O U S
L E V E L L I N G
8 6 . 9 ° C
- 0 . 4 3 b a r 2 9 : 2 1
When the pressure in the sterilization chambers returns within the pre-set safety limits, the
door lock system is released.
As a consequence, the door status indicator
flashes. At the same time, it also beeps.
1
1 3 4 ° C P O R O U S
C Y C L E C O M P L E T E
8 6 . 2 ° C
- 0 . 0 2 b a r 2 9 : 4 0
The icon for the sterilization process status
is steady on.
Standard drying
Intelligent drying
EXTRA DRYING
(+XX) is the time
set