9. RUNNING THE PROGRAM
47
Finally, when the door is opened, you will be asked to reset the device by the following
message:
M A N U A L S T O P
R E S E T S Y S T E M
8 5 . 5 ° C E 9 9 9
- 0 . 0 1 b a r 2 6 : 0 1
To RESET the system, hold down, for at least three seconds, the PROGRAM SELECTION
key until you hear the confirming beep.
When the door is opened, the report for the sterilization cycle executed is produced, including
the error code (E999). Check the report, initial it in the space provided and file it in a suitable
place.
Refer to the print report examples shown in Appendix B, Programs.
After the RESET, the device goes to STAND-BY mode, ready to execute a new program.
NOTE
W
HENEVER AN ALARM IS GENERATED IN CERTAIN PHASES OF THE CYCLE
,
AN
AUTOMATIC PROCEDURE IS ACTIVATED TO CLEAN THE PLUMBING CIRCUIT
.
F
OR A
COMPLETE DESCRIPTION OF THE ALARMS
,
SEE
A
PPENDIX
E
“A
LARMS
”.
WARNING
AFTER A PROGRAM IS MANUALLY INTERRUPTED (MANUAL STOP)
ALWAYS CHECK THE STATUS OF THE ICON
BEFORE USING THE
MATERIAL IN THE STERILIZATION CHAMBER.
IF THE ICON
IS STEADY ON, THE MATERIAL IN THE STERILIZATION
CAN BE CONSIDERED STERILE AND, THUS, BE USED. WE RECOMMEND
USING IT IMMEDIATELY.
HOWEVER, IF IT IS OFF, THE MATERIAL IN THE STERILIZATION
CHAMBER CANNOT BE CONSIDERED STERILE AND ABSOLUTELY MUST
NOT BE USED.