6. CONFIGURATION
30
Setting the printing
mode
(PRINT OPTIONS on the
ADVANCED menu)
Printer model 1
Printer model 2
The sterilizer is equipped with a printer for recording sterilization program data; it is necessary
to set the parameters required for its proper operation.
1. Select PRINT OPTIONS using the
↵
key and the following menu appears:
→
P R I N T E R +
R E P O R T
↑
E X I T
↓
-
Select PRINTER to select the settings for the printer used, or REPORT to set the number
of copies to print and to reprint data from the last program executed.
a) Item
PRINTER
The following screen appears:
→
O F F +
I N T E R N A L
↑
E X T E R N A L
↓
E X I T -
Select OFF to deactivate the printing of data at the end of a sterilization (or test) cycle.
Select INTERNAL to enable the thermal printer set inside the front of the sterilizer. In
this case, another menu opens:
→
T Y P E 1 +
T Y P E 2
↑
E X I T
↓
-
Select Type 1 for the model 1 of the printer installed.
Select Type 2 for the model 2 of the printer installed.
If, on the other hand, you choose EXTERNAL, the data will be printed on an external
peripheral. Following this selection, another menu opens:
→
C R T Y P E +
C R + L F T Y P E
↑
C R ( + F F ) T Y P E
↓
C R + L F ( + F F ) T Y P E -
E X I T
Activate CR to use printers that advance the paper only on the CR (Carriage Return)
command, or CR+LF for that require the CR+LF (Carriage Line Feed)
commands, or with +FF (Form-Feed) for printers that require the addition of this
command.
NOTE
C
ONSULT THE PRINTER MANUAL TO DETERMINE THE TYPE OF COMMAND USED
.
I
F THIS
INFORMATION IS NOT AVAILABLE
,
TRY PRINTING WITH THE VARIOUS OPTIONS TO
IDENTIFY THE CORRECT SETTING
.
Summary of Contents for millenium B2
Page 1: ...Quality System certified by STEAM STERILIZER OPERATING MANUAL Ed 1 Rev 5 Date January 2008...
Page 5: ......
Page 93: ...APPENDIX F DIAGRAMS 88 PLUMBING DIAGRAM Air CHAMBER Recovery tank Air Main tank...
Page 95: ...APPENDIX H NOTES PER THE OPERATOR 90 APPENDIX H NOTES PER THE OPERATOR...