9. RUNNING THE PROGRAM
40
RUNNING
THE CYCLE
INTRODUCTION
A sterilization cycle consists of a determined number of phases. The number and duration of
the phases can differ for the programs, based on the type of air extraction, sterilization process
and drying method.
The electronic control system monitors the various phases, at the same time checking that the
various parameters are respected; if any type of anomaly is encountered during the cycle, the
program is immediately interrupted, generating an alarm identified by a code, with a relative
message explaining the nature of the problem.
With this type of control, when you select a suitable sterilization program, you are guaranteed
perfect sterilization under any conditions.
STARTING THE
CYCLE
Password check
Printer paper-out check
If Millflash is connected
After placing the load in the sterilization chamber (with the precautions explained in the
Chapter, “Preparing the Material”) and selecting the desired program, close the door until
you hear the click.
The door status icon
flashes (door closed).
Press the START button.
NOTE
I
F A PASSWORD HAS BEEN ENABLED WITH THE OPTION
ANY
CYCLE
START
(
SEE THE
C
HAPTER
C
ONFIGURATION
-
S
ETTING THE PASSWORD
),
YOU WILL BE ASKED TO ENTER
THE ACCESS CODE
:
I N S E R T P A S S W O R D
↵
t o e n t e r
Ç
t o e x i t
Enter the password using the + and – keys. Confirm with the
↵
key.
The equipment checks the presence of the paper into the on-board printer; if out or ended the
following message will be displayed:
W A R N I N G
P A P E R O U T
↵
t o c o n t i n u e
Ç
t o e x i t
Push key
↵
to continue however (replace the paper during or at the end of the sterilisation
cycle).
Push key
Ç
to return in Stand-by mode.
Depending on the type of the device connected, the equipment checks the presence of the
Compact Flash card. If not plugged in, the display shows:
W A R N I N G
C F C A R D M I S S I N G
↵
t o c o n t i n u e
Ç
t o e x i t
Plug in the CF card in the recording device and press the key
↵
on the command panel.
Push the key
Ç
to interrupt the start command and return in Stand-by mode.
Summary of Contents for millenium B2