APPENDIX B – PROGRAMS
66
Report following an alarm
Model
MILLENNIUM
B2
S/N
03 BG 0001
Ver. SW
Exxxx/BGyyyyyy
Counter
0007~0015
Selection
134c
POROUS
Temperature
134 °C
Pressure
2.10
Bar
Process time
4 min
Stand-by
HIGH
Pre-vacuum
FRACTIONATED
Drying
STANDARD
CYCLE START
19/11/02
11:30
Time
T1
P
T2
T3
T4
-------------------------------------------------------------------
00:01
CS
075.1 -0.00
130.9
115.2
093.4
00:11
074.9 -0.28
133.3
114.2
094.0
00:21
074.4 -0.46
146.3
113.2
094.5
00:31
074.3 -0.57
152.6
112.2
095.0
00:35 ..
074.3 -0.59
154.2
111.9
095.2
00:51 ..
078.9 -0.62
152.2
110.4
095.6
01:01 ..
074.9 -0.73
146.6 109.6
095.7
01:27 ..
047.8 -0.78
149.3 107.7
095.7
01:57 ..
047.8 -0.80
155.3 105.8
095.4
02:07 ..
076.5 -0.57
149.9
105.2
095.1
02:17 ..
081.1 -0.49
142.1 104.6
094.6
……………
……………
08:15 …
068.4 -0.76
151.8 104.7
102.3
08:22 …
061.1 -0.80
153.6 104.5
101.7
08:32 ...
097.4 +0.01 154.7 104.0
100.8
08:42 …
104.6 +0.24 148.9 103.7
101.0
……………
……………
15:04 …
135.5 +2.15 143.3 111.7
131.7
15:19 …
135.9 +2.17 148.5 113.5
132.6
15:28 …
135.3 +2.16 153.6 115.9
133.0
……………
……………
19:19 …
135.5 +2.15 157.4 126.5
132.5
19:34 …
134.4 +1.07 157.0 126.8
131.2
19:49 …
108.3 +0.25 156.4 126.8
119.9
19:53 DS
104.4 +0.00 156.1 126.6
116.2
STERILISATION NEGATIVE
ALARM CODE:
A112
DESCRIPTION PTC
SHORTCIRCUIT
CAUTION !
PLEASE REFER TO USER MANUAL
----------------------------------------------------------
Cycle Report
HELIX/BD TEST
Model
MILLENNIUM
B2
S/N
03 BG 0001
Ver. SW
Exxxx/BGyyyyyy
Counter
0011/0019
Selection
HELIX/BD
TEST
Temperature
134 °C
Pressure
2.10
bar
Process time
3.5 min
CYCLE START
19/11/02
16:38
Time
C
bar
----------------------------------------------------------
00:01 CS
076.4 +0.00
02:06 1PV 089.3 -0.89
04:35 1PP 120.4 +0.99
05:45 2PV 062.5 -0.78
07:02 2PP 120.2 +0.97
08:15 3PV 061.1 -0.79
11:00 ..
135.6 +2.15
11:14 ..
136.0 +2.17
12:14
135.6 +2.14
13:14
135.6 +2.15
14:14
135.5 +2.14
14:45 ..
135.4 +2.14
15:20 ..
111.5 +0.00
16:34 …
047.8 -0.89
18:21 …
059.5 -0.86
19:21 ..
075.4 -0.50
20:06 CE
078.7 -0.04
12:33 MAX 136.0
14:44 MIN 135.4
Drying pulses
01
CYCLE END
19/11/02
16:38
HELIX/BD TEST COMPLETE
Please attach the indicator hereunder
OPERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
----------------------------------------------------------
Cycle Report
VACUUM TEST
Model
MILLENNIUM
B2
S/N
03 BG 0001
Ver. SW
Exxxx/BGyyyyyy
Counter
0011/0019
Selection
VACUUM
TEST
CYCLE START
19/11/02
11:37
Time
C
bar
----------------------------------------------------------
00:00 CS
035.0 +0.00
01:39 E1F 037.4 -0.80
6:39 E2F 038.4
-0.79
16:39 E3F 042.0 -0.79
17:54 CE
045.5 -0.01
CYCLE END
19/11/02
11:41
VACUUM TEST:
POSITIVE
OPERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
----------------------------------------------------------
Summary of Contents for millenium B2
Page 1: ...Quality System certified by STEAM STERILIZER OPERATING MANUAL Ed 1 Rev 5 Date January 2008...
Page 5: ......
Page 93: ...APPENDIX F DIAGRAMS 88 PLUMBING DIAGRAM Air CHAMBER Recovery tank Air Main tank...
Page 95: ...APPENDIX H NOTES PER THE OPERATOR 90 APPENDIX H NOTES PER THE OPERATOR...