4. INSTALLATION
11
GENERAL
INSTALLATION
PRECAUTIONS
Obey the following warnings for the correct operation of the device and/or to avoid risky
situations:
Install the sterilizer on a flat surface; if necessary, adjust the leveling feet to compensate for
an irregular surface and to slightly tilt lower the front part of the sterilizer.
Make sure that the support surface is strong enough to support the device's weight (about 65
kg);
– Leave adequate space for ventilation (at least 10 cm on each side) all around the
sterilizer, especially in back.
If the device is built-in to a cabinet, be sure to respect the warnings in the preceding
paragraph, avoiding an obstructions to the air intake;
– Do not install the sterilizer near tubs, sinks or similar places, to avoid contact with water or
liquids. This could cause short circuits and/or potentially dangerous situations for the
operator;
– Do not install the sterilizer in a place that is excessively humid or poorly ventilated;
– Do not install the machine were there is gas or inflammable and/or explosive vapors;
– Install the device so that the power cord is not bent or crushed. It must run freely all the
way to the socket.
– Install the device that any external fill/drain tubing is not bent or crushed. They must run
freely to the drain tank.
ELECTRICAL
CONNECTIONS
The sterilizer's must be connected to a socket of the electrical system of adequate capacity for
the device's absorption and ground provided, in conformity with current laws and/or standards.
The socket must be suitably protected by a breaker having the following characteristics:
– Nominal
current
In
16 A
– Differential
current
I
∆
n
0.03 A
WARNING
THE MANUFACTURER WILL NOT BE LIABLE FOR DAMAGES CAUSED BY
INSTALLING THE STERILIZER ON AN INADEQUATE ELECTRICAL
SYSTEM AND/OR NOT EQUIPPED WITH A GROUND.
If it is necessary to replace the plug on the power cord, use one with equal characteristics or,
at any rate, adequate to the device's electrical characteristics. The user is entirely responsible
for the selection and replacement of the plug.
NOTE
A
LWAYS CONNECT THE POWER CORD DIRECTLY TO THE SOCKET
.
D
O NOT USE
EXTENSION CORDS
,
ADAPTERS OR OTHER ACCESSORIES
.
CONNECTING THE
DATA RECORDING
MILLFLASH
The sterilizer can be connected to MILLFLASH allowing the recording of the cycle data on .txt
format file and its management by PC.
The connectors of the service box are used for interfacing; refer to MILLFLASH Operating
Manual for the installation instructions.
Summary of Contents for millenium B2
Page 1: ...Quality System certified by STEAM STERILIZER OPERATING MANUAL Ed 1 Rev 5 Date January 2008...
Page 5: ......
Page 93: ...APPENDIX F DIAGRAMS 88 PLUMBING DIAGRAM Air CHAMBER Recovery tank Air Main tank...
Page 95: ...APPENDIX H NOTES PER THE OPERATOR 90 APPENDIX H NOTES PER THE OPERATOR...