Operating instructions
CLIMACELL
Page 7
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
3.1.1 Useful
space
Useful space - see the illustration. in the following picture, where X(H) = X(D) = 10 % of the
chamber depth, X(Š) = X(W) = 10 % of the chamber width, X(V)
1
= X(H) = distance of the
bottom shelf to the bottom of the chamber. The required temperature accuracy from the par.
Technical parameters can be only achieved within the space defined above(according to DIN
12 880 – marked with heavy lines in the picture, the light lines mark the inner walls of the
chamber). (It means that over the highest shelf the values of the chapter Technical
parameters are not obligatory).
3.2 Connector
RS-232C
for protocol printer
Interface parameters
Baud: 4800
Stopbit: 1
Parity: none
Databit: 8
Devices connected with of connector RS 232 must fulfill valid regulations and be
approved by a competent testing laboratory.
The unit is designed for connecting to printer CITIZEN, model iDP 3110-24 RF-A, which
can be ordered optionally.
Pin
Signal
2
TxD
3
RxD
5
GND
6
DSR