Operating instructions CLIMACELL Page 30
V 1.07_Blue Line CLIMACELL_PCH_en_np_mmm
!
0208
Error 06 = steam generator overfilled (upper level indicator flooded for a longer time than
three minutes)
∗
Error 07 = water supply interrupted (the filling valve has been opened for a longer time than
90 sec – drinking water, 180 sec – distilled water and the upper level indicator has not been
flooded yet)
∗
Error 08 – 11 = device failure –used for developing purposes
Error 12 = the third PT 100 sensor (in the steam generator) not connected
∗
Error 13 = pump failure (it passed 120+60 sec since the pump was started five times and
the upper level indicator has not been deflooded yet)
∗
Error 14 = level indicators failure (lower level indicator reports deflooding, upper level
indicator reports flooding)
∗
Error 15 =. the RH sensor failure (or a false calibration during the service)
∗
Error 16 =. repeated error 13 in repeated test
∗
(Note: the pump pumps only in case of full flooding, when the water level reaches the
switching-off level of the upper level indicator swimmer – see the swimmer positions in the
picture below)
* It is a soft error. The difference between the soft error and the hard one is, that after the soft
error occurs and is reported the whole running process is not stopped, only the system is
eliminated, whose element has been detected as defective, in case of Climacell it is steam
generator. Other systems continue their controlled function – heating, cooling, ventilator,
lighting. If such an error occurs – the regulator (without interrupting the operation) informs –
acoustically and optically by means of display – the operator. The acoustic signal is the same
as in case of the hard error, the optical signal differs (from reporting the hard error) by
flickering of the appropriate message on the display. In case of hard error it is shown
permanently.
The soft error is used, because its cause can be removed by a simple correction and it is not
necessary to interrupt the running process. E.g. the hose of the drain from the steam
generator is rolled up improperly, which prevents the water to go out when turbidity is
removed, on the other hand it is not necessary to interrupt the running process by a hard
error.
If a “soft” error occurs, you can proceed in following ways:
1) By depressing X/W only the acoustic signal is suppressed so that it does not disturb. The
humidity generating elements are blocked, the unit does not regulate the humidity. Other
functions are kept and controlled.
2) By depressing STOP
- both the acoustic signal is suppressed
- and the humidity generating elements are unblocked. During the “soft” error the button
STOP has not the function of stopping the operation. After depressing STOP, if the soft
error cause is eliminated, the unit can continue the operation inclusive the humidity. If
another error occurs in future, it will be signalized only optically, not acoustically.
3) By disconnecting the unit from mains and connecting it again (e.g. with the main switch)
during the operation the elements of the humidity automatics are initialized and the unit
starts running automatically in the same way as in case of the fall-out loss of electrical
energy and the operation continues from the point of interruption. In this case another
eventual “soft” error will be signalized both acoustically and optically.
If “soft” errors occur repeatedly, please inform your service organization.