After the set temperature has been reached, the run down time is shown if the time of the
procedure is unlimited. If the time is limited, the rest of the run down time is shown.
4.7.4 Information on the set parameters during the device’s running:
See paragraph 4.2.4
4.7.5 Stop P3
Press START/STOP.
4.8 Program
P4
With two temperature levels T1 and T2, with controlled rate of temperature variation (ramps),
with setting the RH for each temperature level and with cycling option.
Operates at 2 temperature levels T1 and T2, the dwell time on the level T1 is limited, on the
level T2 can be limited or unlimited.
4.8.1 Set
P4
Index: T0 = initial chamber temperature
T1 = temperature level 1
T2 = temperature level 2
Digit 11 on LCD, placed between temperature and dwell time values, shows the
following information:
r = ramp value, can be set on digits 12 – 16
and corresponds with temperature change speed in °C/min
max. value is +3.20 °C/min, when T0 < T1
max. value is –1.0 °C/min, when T0 > T1
If r = 00.0, no ramp active!
(Variations of maximum ramp rates by different temperatures may appear due to physical
limits of heating power.)
t = time available in position
"
(00h:00min – 99 h 59 min), at T2 the dwell time can
be unlimited
c = cycles shown in position
"
, digits 12 – 13 on LCD (available only at T2, if the
dwell time is unlimited)
Possible number of cycles is 02 to 99 (eventually
∞
- which is the next position after 99 - the
cycling is possible on condition, that the temperature drops below the value T0 after the end
of interval on T2)