Operating instructions
CLIMACELL
Page 15
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
4.5 Program
P1
A constant temperature and a constant value of RH is required.
P 1 =
The cabinet is working fluently at the set temperature and RH until the unit is switched off.
4.5.1 Set
P1
Put cursor to the position Program with help of
◄►
and switch to P1 with help of
▲
Set the temperature at the position °C between 0 and 99.9 °C in steps 0.1
°
C Place the
cursor on the lower line of the display and set the ventilator rate at the position
to a value
from 00 % to 100 %in 10 %-steps (% of maximum rate) and RH between 10 % and 90 %
with limitations given in diagram in paragraph
Setting the operating conditions of
temperature and relative humidity and their limitations.
CAUTION!
Do not reduce the ventilator rate if not necessary. The reduced ventilator rate could
influence the regulation of temperature and RH negatively!
4.5.2 LCD read out shown
4.5.3 Start P1
Press START/STOP, a sound will appear and the “windmill“ in the first digit of LCD starts
running clockwise.
See the following LCD read out:
Digit 1 (windmill running) = program running
Digit 7 to 10 = actual temperature in °C.
Digit 12 to 16 = time from the start in hr:min.
As soon as the set values are reached, then run down time is shown.