4.8.2 Set the segment for T1
4.8.2.1 Set temperature T1
Put cursor to the position Program and switch to P4. By pressing
◄►
place the cursor on
the graph and make sure that T1 (
/
¯
\
_ ) is activated. If not, activate it with button
▲
.
Change to the position °C - setting the temperature - and set the temperature between.
4.8.2.2 Set the delay time of T1 :
Move cursor to the digit 11 on LCD and ask for parameters by pressing
▲
Either „r“ or „t“ will
appear (see above). Select „t“. Move to the position
"
and set the dwell time between 00
hr:00 min and 99 hr 59 min.
4.8.2.3 Read out of T1 and the delay time:
4.8.2.4 Set the ramp at T1 :
Return the cursor to the digit 11 on LCD and by pressing
▲
select „r“. Switch to the position
"
and set the required ramp (max. value +3.20 °C/min for T0
<
T1). If the ramp is not
needed, leave „r“ on 0.00. In this case the temperature will be changed at maximum speed.
The symbol
±
appears automatically in accordance to T0 and T1 relation.
4.8.2.5 Read out of T1 and the ramp:
4.8.2.6 Set the fan speed at the temperature T1:
Place the cursor at the position
and set the ventilator rate between 00 to 100 %. Then set
the required relative humidity between 10 % and 90 % at the position RH – with limitations
given in diagram in paragraph
Setting the operating conditions of temperature and
relative humidity and their limitations.
Caution!
Do not reduce the ventilator rate if not necessary. The reduced ventilator rate could
influence the regulation of temperature and RH negatively!