Operating instructions
CLIMACELL
Page 11
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
The procedure is as follows: By pressing START/STOP you enter the program restoring
service at first. The choices P1 – P6 are available. Select by pressing
▲
. To record the
program press button
.
(In detail:
Before the start:
the display indicates: „Service 02“
„Load from CCard“
After the start:
on the second line of the display by pressing
▲
select what will loaded from
the chip card. Choices:
- all eligible parameters of P1 to P4 at once - the display indicates:
„MMM CCard
→
P1
÷
4“,
- all eligible parameters of P5 - the display indicates:
„MMM CCard
→
P5“,
- all eligible parameters of P6 - the display indicates: „MMM CCard
→
P6“.
The loading of your choice from the chip card is done by pressing
If an incorrect card is
placed into the unit, the display indicates: “Bad MMM CCard“. If everything is all right, the
indication is: “Done“.)
Quit the service by pressing START/STOP again.
Caution!
Use only an original chip card delivered by the producer. Using of another card could
cause damaging of your unit’s regulator and the guarantee would be cancelled.
4.4.4 Service 03 – Set the real time and date
Pressing START/STOP you enter the service. Set the real time and date by pressing buttons
◄►
and
▲
. To quit the service press START/STOP again. Data on the display are in the
following form: ho:mi_day:month:year
4.4.5 Service 04 – Set interval on printer/communication interface
To print protocols device can be set up (optional) with a printer CITIZEN iDP 3110-24 RF-A,
which is connected by a cable to the printer interface – see par. Printing the protocol. Enter
the service by START/STOP and set the interval:
Pressing
▲
select one of the following choices:
printer off
switched over to PC (application of SW DS Com – data are sent to PC)
interval 10 s
interval 01 min
interval 01 h
By pressing
◄
move the cursor to the position 10/01 seconds, minutes or hours. By pressing
▲
you can change the printer interval within the following range:
10 to 50 s (step 10 s)
01 to 59 min (step 1 min)
01 to 12 h (step 1 h)
To quit the service press START/STOP again.
.
Caution!
Only one unit is active, either seconds, minutes, or hours!