Operating instructions CLIMACELL Page 24
V 1.07_Blue Line CLIMACELL_PCH_en_np_mmm
!
0208
4.8.6 Stop
P4
Press START/STOP.
4.9
Program P5, P6
4.9.1 Description of P5 and its control
The program consists of 16 segments. Segment 0 is for the delayed start, it is defined only
by the time t0. If t0 is 0, there is no delayed start. Segments 1 to 15 are of the same qualities,
each of them is defined only by the end temperature, time or speed (ramp), fan rotation
speed and depending on the type of the unit also by relative humidity and chamber light.
Each of the segments
S
1 to
S
15 can be set as the last and the infinite, that is after the end
temperature
T
i has been reached, and simultaneously after the time
t
i is over either the
program is finished or followed by another cycle.
Each of the segments 1 to 15 can be set as the last and infinite, that is the program at this
segment must be finished by START/STOP button.
If the cycle selection is active on LCD at the moment of the start, see fig. Display - state of
the program selection - E: cXX, the program comes again to the start after the last segment
is over and the whole program is repeated. The definite number of program cycles
corresponds with the number behind the letter “c“. The segment 0 can be counted into the
cycle as well even when it is zero.
The rise to the set RH starts after reaching the set
temperature.
Range of parameters:
time
t0
0 hr, 0min to 99hr 59 min
time
t1 to t15
0 hr, 0 min. to 99 hr, 59 min., or infinite
ramp
r1 to r15
0 to 3.2 °C/min for heating
-1.0 to 0 °C for cooling
temperature
T1 to T15
0,0 to 99,9 °C
fan speed
V1 to V15
0 to100% with 10% steps
relative humidity
RH1 to RH15
10 % to 90 % with 1 %
light
L1 to L15
ON/OFF (only for the type CLC-L)
number of cycles C
2 to 99 (eventually
∞
- which is the next
position after 99)