Operating instructions CLIMACELL Page 26
V 1.07_Blue Line CLIMACELL_PCH_en_np_mmm
!
0208
Display - running of the program
A B C D E
F G H
A:
∗
5
the rotating symbol and the program number
B: XX
the number of the relevant segment
C: X
the type of the set segment:
> not the last, the end length - see E
>I the last, the end length - see E
∞
the last, infinite length
D: XX . X
the temperature in the chamber at the moment
E: XX : XX
the segment time at the moment - hours, minutes
all segments - counting down from the set value towards zero,
in case of an infinite length of the last segment counting up from zero
in case the time is higher than 99 hr. 59 min >>:>> is indicated
F: XXX the fan speed at the moment
G: XX
disengaged
H: XXX
the light at the moment
4.9.2 Description of P6 and its control
The program consists of 16 segments. Segment 0 represents delayed start, it is determined
only by the time t0. If a sign “x” is in the display C, then there is no delayed start. Segments 1
to 15 have the same functions, each of them is determined by the end temperature, time or
rise rate (ramp), ventilator rate, relative humidity RH and according to the unit type also by
the exposure lighting of the chamber.
Each of the segments
S
1 to
S
15 can be set as the last, that is after the end temperature
T
i
has been reached, and simultaneously after the time
t
i is over either the program is finished
or followed by another cycle.
Each of the segments 1 to 15 can be set as the last and infinite, that is the program at this
segment must be finished by START/STOP button.