Operating instructions CLIMACELL Page 4
V 1.07_Blue Line CLIMACELL_PCH_en_np_mmm
!
0208
2.2
Before putting into operation
CLIMACELL is designed and produced in accordance with EN 50081-1, 55011, 50082-2,
61000-4-2, 61000-4-4, ENV 50140, 50141, meets requirements of EU directions
89/336/EEC, 73/23/EEC, and is individually tested according to EN 61010-1.
♦
Before starting the work with this unit please study the instructions for use
carefully!
♦
Lay the treated material only on the sieves in the unit, never on the bottom of the
chamber!
♦
Carrying capacity of the floor, where the unit will be placed, must correspond to the
weight of the unit itself plus the weight of the maximum charge (see chapter
Technical
parameters
).
♦
The unit must not be placed on a floor covering, which could cause a danger of fire
or smothering in case of falling of hot objects out of the unit.
♦
No inflammable, explosive or toxic stuff or materials which could give such
stuff off may be inserted into the units!
♦
Unit is not designed for warming of liquids.
♦
Units are not designed for using in atmosphere with danger of inflammable or
explosive anaesthetics.
♦
Every mounting and demounting of the unit’s parts may
be carried out only
after disconnecting the unit from the mains by pulling the supply cord out of the
socket.
♦
Installation of the unit is carried out by connecting to the mains and connecting to the
water source and the drain. Parameters of the electric connections are specified in
chapter
Electric installation and other conditions.
♦
If the unit is not used for a longer time, disconnect it from the mains by pulling the
connecting cord out of the socket.
♦
The mains cord must not come into contact with hot parts of the unit
♦
Protection of the temperature cabinet, its surroundings and the treated material from
surpassing the temperature is secured by a safety thermostat. The unit
CLIMACELL has
a safety thermostat class 3 according to EN 61010-2-010.
♦
Check the function of the safety thermostat regularly – every day (see paragraph
Check of the function of the safety thermostat).
♦
The types of cabinets are: standing on the floor, space saving, equipped with rolls
and a brake. It is necessary, that the cooling incubator stands on an even floor.
Otherwise the unit’s run could be influenced negatively.
♦
Minimum distance of the rear and side wall of the unit from other objects and walls
is 100 mm, from the upper wall it is 400 mm.