Troubleshooting Chart
Chapter Eight: Troubleshooting
85
MFC output signal
matches setpoint, but
actual gas flow less (as
determined by transfer
standard)
Contaminated bypass
Check process chamber pressure.
Compare to normal or reference.
MFC programmed for different
gas
Compare gas programmed in MFC to
actual gas used. Verify using DNet
or Ethernet interface.
MFC cannot achieve FS
flow
Inlet pressure low
Increase inlet pressure
Outlet pressure high
Decrease outlet pressure
Valve contamination
Check valve current for maximum
position
Gas
line
blockage/contamination
Measure pressure drop across
component suspected of
contamination such as filter or check
valve
MFC setpoint in counts > than
100% FS of 24567 counts
Program attributed 6 to 100% =
24567 counts (0x6000)
Output signal matches
setpoint @ higher flows,
but will not go to 0
MFC valve partial contamination Cycle-purge device, check valve
current
Faulty control valve, adjustment
or electronics
Run MFC diagnostics, call MKS
service center
Inlet pressure too high
Decrease inlet pressure
Device zero offset
re-zero device with known zero flow
conditions
Display powers up, but
MFC does not respond
Bad DNet/RS485 connection
Check status DNet polled
connection/RS485 connection. check
network LED
Incorrect MAC ID Address
Check MAC ID on device
Incorrect baud rate setting
Check baud rate on device
Incorrect DNet I/O Instance
Setting
Check I/O setting using DNet
commissioning tool/software
Check tool host
Low power, power supply
Measure current & voltage from
power source
Summary of Contents for P9B
Page 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Page 4: ......
Page 10: ...List of References vi This page intentionally left blank...
Page 20: ......
Page 22: ...12...
Page 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Page 24: ...14...
Page 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Page 27: ...17...
Page 28: ...18...
Page 32: ......
Page 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Page 64: ......
Page 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Page 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Page 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Page 100: ......
Page 105: ...Appendix A Product Specifications Electrical Specifications 95...
Page 111: ...Model Code Description 101...
Page 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...