Appendix A: Product Specifications Performance
Specifications
–P9B
91
Appendix A: Product Specifications
Performance Specifications –P9B
Full Scale Flow (N2 equivalent) 5-50000
sccm
Maximum Inlet Pressure
150 psig, limited to maximum differential pressure across MFC.
Normal Operating Pressure Differential (with atmospheric
pressure at the MFC outlet)
10 to 5000 sccm, 10 to 40 psid
10000 to 30000 sccm 15 to 40 psid
50000 sccm 25 to 40 psid
Proof Pressure
1000 psig
Burst Pressure
1500 psig
Control Range
2% to 100% of F.S.
Flow Accuracy
1% of setpoint for > 20 to 100% F.S.
0.2% of FS for 2 to 20% F.S.
Flow Repeatability
0.3% of Reading
Flow Resolution
0.1% of Full Scale
Temperature Coefficients
Zero
Span
<0.05% F.S./
C
<0.08% Reading/
C
Inlet Pressure Coefficient
< 0.02% of Reading./psi
Pressure Sensitivity
Less than 5% flow perturbation from 20 to 100% of FS when
subject to a pressure transient of 2 psi
Typical Controller Settling Time (per SEMI Guideline E17-
0600)
~ 750 milliseconds typical above 5% F.S.
Pressure Reading
1 to 100 psia +/- 1% of Full Scale of 100 psia
Warm-Up Time
(to within 0.2% of F.S. of steady state performance)
< 30 min.
Normal Operating Temperature Range
10
C to 50
C
Storage Humidity
0 to 95% Relative Humidity, non-condensing
Storage Temperature
-20
C to 65
C (-4
F to 149
F)
Temperature Display Range
0 to 100
C
Temperature Readout Units
C
Temperature Accuracy
2
C
Temperature Resolution
0.1
C
ROHS Compliant
Yes
Specifications are subject to change without notice.
Summary of Contents for P9B
Page 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Page 4: ......
Page 10: ...List of References vi This page intentionally left blank...
Page 20: ......
Page 22: ...12...
Page 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Page 24: ...14...
Page 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Page 27: ...17...
Page 28: ...18...
Page 32: ......
Page 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Page 64: ......
Page 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Page 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Page 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Page 100: ......
Page 105: ...Appendix A Product Specifications Electrical Specifications 95...
Page 111: ...Model Code Description 101...
Page 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...