Tuning the MFC for Optional PC
Chapter Six: Embedded Web-Based GUI and Diagnostics
75
Adjusting the Proportional and Integral Control and Optional Command Min Offset
Over Ethernet you have the ability to view the current values for the proportional control setting (Kp), the
integral control setting (Ki) and the Optional Command Min Offset, and change them. To do this, follow the
steps listed below:
1. Connect to the Embedded GUI, see
Logging On to Your MFC
on page 58.
2. Enter Setup Mode, see
Setup Mode
on page 64.
3. Go to the Plot Page. Here you are able to adjust the settings. See the Setup Mode
Plot Page
section on
page 67 for further instructions.
Summary of Contents for P9B
Page 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Page 4: ......
Page 10: ...List of References vi This page intentionally left blank...
Page 20: ......
Page 22: ...12...
Page 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Page 24: ...14...
Page 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Page 27: ...17...
Page 28: ...18...
Page 32: ......
Page 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Page 64: ......
Page 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Page 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Page 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Page 100: ......
Page 105: ...Appendix A Product Specifications Electrical Specifications 95...
Page 111: ...Model Code Description 101...
Page 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...