Measurement Limitations
Monitoring NIBP
6 - 4
BP10 Operator’s Manual
6.3
Measurement Limitations
Measurements are impossible with heart rate extremes of less than 30 bpm or greater
than 300 bpm, or if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible:
■
If a regular arterial pressure pulse is hard to detect.
■
With excessive and continuous patient movement such as shivering or
convulsions.
■
With cardiac arrhythmias.
■
Rapid blood pressure changes.
■
Severe shock or hypothermia that reduces blood flow to the peripheries.
NOTE
•
The effectiveness of this sphygmomanometer has not been established in
pregnant, including pre-eclamptic patients.
6.4
Measuring NIBP
6.4.1
Preparing the Patient
In order to minimize the NIBP measurement errors, whenever possible make sure the
patient is in the following positions.
■
Comfortably seated
■
Legs uncrossed
■
Feet flat on the floor
■
Back and arm supported
■
Middle of the cuff at the level of the right atrium of the heart
NOTE
•
It is recommended that the patient relaxes as much as possible and keep
calm before performing measurement and that the patient does not talk
during NIBP measurement.
•
It is recommended that 5 minutes should elapse before the first
measurement is taken.
•
The operator should not touch the cuff or tubing during NIBP measurement.
Summary of Contents for BP10 NIBP Module
Page 1: ...BP10 NIBP Module Operator s Manual...
Page 2: ......
Page 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Page 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Page 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Page 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Page 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Page 87: ...BP10 Operator s Manual B 1 B EMC...
Page 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Page 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Page 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Page 99: ......
Page 100: ...P N 046 011008 00 4 0...