Monitoring NIBP
Configuring NIBP Measurement Modes
BP10 Operator’s Manual
6 - 7
6.6
Configuring NIBP Measurement Modes
BP10 provides five measurement modes:
■
Manual mode
■
Continuous mode
■
Automatic mode
■
Sequence mode
■
ABPM (Ambulatory Blood Pressure Monitoring) mode
The automatic, sequence and ABPM modes need to configure more settings.
NOTE
•
You can initiate a manual or continuous measurement in the course of
automatic, sequence and ABPM series measurements.
6.6.1
Setting the Automatic Mode
In automatic mode, the NIBP measurements are automatically taken on the configured
interval.
To enable and set up the automatic mode, follow this procedure:
1.
In the main menu, select
Mode
and set it to
Auto
.
2.
Select
Interval
and select the appropriate period. The default setting is 15 min.
6.6.2
Setting the Sequence Mode
In the sequence mode, the NIBP measurements are automatically taken in the user-
defined sequence, which allows you to set different measurement intervals in
consecutive five (5) phases.
To enable and set up the sequence mode, follow this procedure:
1.
In the main menu, select
Mode
and set it to
Sequence
.
2.
Select
Sequence Settings
and open the
Sequence Settings
screen.
3.
Set up the interval and duration in each phase.
Options
Description
Settings
Phase
In a sequence, you can set up to five
measurement phases from P1 to P5.
Note:
If some existing phases need
to be removed, set the phase
interval to OFF.
/
Summary of Contents for BP10 NIBP Module
Page 1: ...BP10 NIBP Module Operator s Manual...
Page 2: ......
Page 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Page 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Page 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Page 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Page 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Page 87: ...BP10 Operator s Manual B 1 B EMC...
Page 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Page 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Page 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Page 99: ......
Page 100: ...P N 046 011008 00 4 0...