Configuring NIBP Measurement Modes
Monitoring NIBP
6 - 8
BP10 Operator’s Manual
6.6.3
Setting up the ABPM Mode
In ABPM mode, the NIBP measurements are automatically taken according to the
configured intervals for the day and the night.
To enable and set up the ABPM mode, follow this procedure:
1.
In the
Menu
screen, select
Mode
and set it to
ABPM
.
2.
Select
ABPM Settings
and open the
ABPM Settings
screen.
Set the start time and interval for the day and night respectively.
The default start time of the day is 6:00, and the interval is 15 min.
The default start time of the night is 22:00, and the interval is 30 min.
6.6.4
About Continuous Mode
In the continuous mode, the measurements are taken one after another in a five-minute
period. After that, the continuous mode ends, and the device restores to the NIBP mode
before the continuous mode starts.
To enable the continuous mode, enter the main menu and select
NIBP STAT
.
Interval
Set up the interval between NIBP
measurements in an individual phase.
1 min, 2 min, 2.5 min, 3 min,
5 min, 10 min, 15 min, 20
min, 30 min, 1 h, 1.5 h, 2 h,
Off.
Duration
Set up the lasting time of an individual
phase.
10 min, 30 min, 1 h, 2 h, 4 h,
8 h, continuous
Options
Description
Settings
Summary of Contents for BP10 NIBP Module
Page 1: ...BP10 NIBP Module Operator s Manual...
Page 2: ......
Page 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Page 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Page 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Page 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Page 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Page 87: ...BP10 Operator s Manual B 1 B EMC...
Page 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Page 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Page 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Page 99: ......
Page 100: ...P N 046 011008 00 4 0...