General Problems
Troubleshooting
12 - 2
BP10 Operator’s Manual
12.1
General Problems
12.2
Technical Messages and Solutions
The following table lists the major technical messages displayed in the message area of
BP10.
For the clearable technical messages, you can clear them by selecting
Clear Prompt
in the
main menu.
Symptom
Possible Cause
Solution
Cannot discharge patient, change
patient category, or set NIBP
measurement related configurations
at BP10.
BP10 is paired with TD60/TM80.
Unpair BP10 from TD60/TM80 by
pressing the
key, and then
discharge the patient by
selecting
Discharge Patient
in
the main menu at BP10.
The BP10 cannot be paired with the
TD60/TM80.
1. The distance between BP10
and TD60/TM80 is out of the
radio range.
2. Signal is interrupted.
3. The TD60/TM80 is already
paired with other BP10.
4. After BP10 is paired with TD60/
TM80, user turns off and then
turns on the MPAN at BP10 or
TD60/TM80.
Restart the pairing atTD60/
TM80.
Message
Clearable
(Yes/No)?
Possible cause
Solution
Low Battery
No
The battery charge is low.
Replace with a new battery.
Battery Depleted
No
The battery charge is almost
depleted.
Voltage Error
No
The lithium-ion battery voltage is
out of the specified range.
Battery Error
No
The lithium-ion battery
communication is error.
Verify that the battery is
correctly installed.
If the error persists, replace with
a new battery.
Battery Maintenance
Required
No
The lithium-ion battery is aging.
Replace with a new battery.
MPAN Disconnected
Yes
1. The distance between BP10
and TD60/TM80 is out of the
radio range.
2. The TD60/TM80 is powered off.
1. Move the BP10 and TD60/
TM80 closely.
2. Power on the TD60/TM80.
Summary of Contents for BP10 NIBP Module
Page 1: ...BP10 NIBP Module Operator s Manual...
Page 2: ......
Page 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Page 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Page 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Page 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Page 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Page 87: ...BP10 Operator s Manual B 1 B EMC...
Page 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Page 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Page 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Page 99: ......
Page 100: ...P N 046 011008 00 4 0...