Monitoring ECG
Preparation for Monitoring ECG
TM80/TM70 Telemetry Monitor Operator’s Manual
7 - 11
When any of the leads are off, the indications are as follows:
■
The lead off message displays on the information bar.
The background color of the information bar corresponds to the alarm level.
■
A flashing circle indicates the disconnected lead.
The color of the flashing circle is based on the alarm level.
7.3.6
Checking the Paced Status
It is important to correctly set the patient’s paced status before you start monitoring
ECG.
To check the paced status:
1.
On the main screen, tap the HR digital area or ECG waveform area to enter the
[
ECG
] menu.
2.
In the [
Pacer
] section, check the setting of the paced status.
The current paced status setting displays to the right of [
Paced
].
3.
If the paced status setting is not correct, tap [
Paced
] and select the correct paced
status.
You can also change the patient’s paced status from the [
Patient Info
] menu. For more
"Changing Paced Status" on page 5 - 3
.
■
When [
Paced
] is set to [
Yes
] at the TM80/TM70, if the pacer pulse is detected, the
symbol displays in the waveform area of the CMS’ screen, and the pace pulse
marks will display on the ECG waveform both at the TM80/TM70 and CMS.
■
When [
Paced
] is set to [
No
] or not specified at the TM80/TM70, the
symbol
displays in the waveform area of the CMS’ screen.